The Effect of Settling on the Resonance Frequency Analysis in Intermediate Components on Implants: an In Vitro Study

Authors

  • Lara Rúbia Marques Nascimento Graduanda em Odontologia, Departamento de Odontologia Restauradora, Faculdade de Odontologia, UFMG Univ. Federal de Minas Gerais, 31270-901 Belo Horizonte - MG, Brasil https://orcid.org/0009-0004-0483-5600
  • Clara Almeida Mares Cirurgiã-Dentista; Departamento de Odontologia Restauradora, Faculdade de Odontologia, UFMG Univ. Federal de Minas Gerais, 31270-901 Belo Horizonte - MG, Brasil https://orcid.org/0000-0002-8592-0003
  • Fernando Teodoro Metzker Lyra Cirurgião-Dentista; Departamento de Odontologia Restauradora, Faculdade de Odontologia, UFMG Univ. Federal de Minas Gerais, 31270-901 Belo Horizonte - MG, Brasil https://orcid.org/0000-0002-3898-8339
  • Rafaela Regina de Lima Doutoranda, Programa de Pós-Graduação em Odontologia, Faculdade de Odontologia, UFMG Univ. Federal de Minas Gerais, 31270-901 Belo Horizonte - MG, Brasil https://orcid.org/0009-0000-6008-4347
  • Danilo Rocha Dias Professor Doutor, Departamento de Odontologia Restauradora, Faculdade de Odontologia, UFMG Univ. Federal de Minas Gerais, 31270-901 Belo Horizonte - MG, Brasil https://orcid.org/0000-0002-9194-8774
  • Frederico Santos Lages Professor Doutor, Departamento de Odontologia Restauradora, Faculdade de Odontologia, UFMG Univ. Federal de Minas Gerais, 31270-901 Belo Horizonte - MG, Brasil https://orcid.org/0000-0002-8885-1138

DOI:

https://doi.org/10.21270/archi.v14i1.6446

Keywords:

Resonance Frequency Analysis, Biomechanics, Torque, Dental Implants

Abstract

Introduction: When applying torque to the screw of the prosthetic component, a compressive force develops between the parts, and the screw-implant connection is subject to the effect of settling. The interior of the screw and implant threads are manufactured with irregularities that are smoothed out by the tightening force, causing a loss of 2-10% of the initial torque. Therefore, it is recommended to reapply the torque after 10 minutes. Resonance frequency analysis is the main method for evaluating the long-term stability of implants, and its behavior on intermediate components still raises questions. Objectives: To evaluate whether the effect of settling significantly interferes with the resonance frequency analysis values on the screw of the component. Methodology: Alternating installation of 10 Cone Morse and 10 External Hexagon dental implants in blocks of fresh bovine ribs, selected according to the same bone density. Intermediate components of the mini conical abutment type were installed with torque values of 20 N.cm and 32 N.cm. The Ostell device was used to measure the resonance frequency analysis on the 4 sides of the components, and the average values were considered. After 10 minutes, new measurements were taken. Results: The immediate analysis values on the dental implant components and after 10 minutes were not significantly different. Conclusion: Within the limitations of the study design, it was found that the effect of settling does not interfere with the resonance frequency analysis of the implant's intermediate components.

Downloads

Download data is not yet available.

References

1. Adell R, Lekholm U, Rockler B, Brånemark PI. A 15-year study of osseointegrated implants in the treatment of the edentulous jaw. Int J Oral Surg. 1981;10(6):387-416.

2. Huang Y, Wang J. Mechanism of and factors associated with the loosening of the implant abutment screw: A review. J Esthet Restor Dent. 2019;31(4):338-345

3. Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol. 2002;28(2):67-73.

4. Dixon DL, Breeding LC, Sadler JP, McKay ML. Comparison of screw loosening, rotation, and deflection among three implant designs. J Prosthet Dent. 1995;74(3):270-8.

5. Xiao JR, Li YQ, Guan SM, Kong L, Liu B, Li D. Effects of lateral cortical anchorage on the primary stability of implants subjected to controlled loads: an in vitro study. Br J Oral Maxillofac Surg. 2012;50(2):161-5.

6. Manzano-Moreno FJ, Herrera-Briones FJ, Bassam T, Vallecillo-Capilla MF, Reyes-Botella C. Factors Affecting Dental Implant Stability Measured Using the Ostell Mentor Device: A Systematic Review. Implant Dent. 2015;24(5):565-77.

7. Lages FS, Douglas-de Oliveira DW, Costa FO. Relationship between implant stability measurements obtained by insertion torque and resonance frequency analysis: A systematic review. Clin Implant Dent Relat Res. 2018;20(1):26-33.

8. Sennerby L, Meredith N. Implant stability measurements using resonance frequency analysis: biological and biomechanical aspects and clinical implications. Periodontol 2000. 2008;47:51-66.

9. Breeding LC, Dixon DL, Nelson EW, Tietge JD. Torque required to loosen single-tooth implant abutment screws before and after simulated function. Int J Prosthodont. 1993;6(5):435-9.

10. Sakaguchi RL, Borgersen SE. Nonlinear contact analysis of preload in dental implant screws. Int J Oral Maxillofac Implants. 1995;10(3):295-302.

11. Bakaeen LG, Winkler S, Neff PA. The effect of implant diameter, restoration design, and occlusal table variations on screw loosening of posterior single-tooth implant restorations. J Oral Implantol. 2001;27(2):63-72.

12. de Molon RS, Lages FS, Rivera CP, de Souza Faloni AP, Margonar R, Queiroz TP. Evaluation of Short and Regular Implants after Prosthesis Placement in the Mandible: A Nonrandomized Controlled Clinical Trial. J Contemp Dent Pract. 2017;18(12):1122-1129

13. López-Jarana P, Díaz-Castro CM, Falcão A, Ríos-Carrasco B, Fernandez-Palacín A, Ríos-Santos JV, Herrero-Climent M. Is It Possible to Monitor Implant Stability on a Prosthetic Abutment? An In Vitro Resonance Frequency Analysis. Int J Environ Res Public Health. 2020;17(11):4073.

14. Lages FS, Willya Douglas-de-Oliveira D, Ibelli GS, Assaf F, Queiroz TP, Costa FO. Relationship between implant stability on the abutment and platform level by means of resonance frequency analysis: A cross-sectional study. PLoS One. 2017;12(7):e0181873.

15. Winkler S, Ring K, Ring JD, Boberick KG. Implant screw mechanics and the settling effect: overview. J Oral Implantol. 2003;29(5):242-5.

16. Byrne D, Jacobs S, O'Connell B, Houston F, Claffey N. Preloads generated with repeated tightening in three types of screws used in dental implant assemblies. J Prosthodont. 2006;15(3):164-71.

17. Lee HW, Alkumru H, Ganss B, Lai JY, Ramp LC, Liu PR. The Effect of Contamination of Implant Screws on Reverse Torque. Int J Oral Maxillofac Implants. 2015;30(5):1054-60.

18. Calcaterra R, Di Girolamo M, Mirisola C, Baggi L. Effects of Repeated Screw Tightening on Implant Abutment Interfaces in Terms of Bacterial and Yeast Leakage in Vitro: One-Time Abutment Versus the Multiscrewing Technique. Int J Periodontics Restorative Dent. 2016;36(2):275-80.

19. Molina A, Sanz-Sánchez I, Martín C, Blanco J, Sanz M. The effect of one-time abutment placement on interproximal bone levels and peri-implant soft tissues: a prospective randomized clinical trial. Clin Oral Implants Res. 2017;28(4):443-452.

Published

2025-01-31

How to Cite

Nascimento, L. R. M., Mares, C. A., Lyra, F. T. M., Lima, R. R. de, Dias, D. R., & Lages, F. S. (2025). The Effect of Settling on the Resonance Frequency Analysis in Intermediate Components on Implants: an In Vitro Study. ARCHIVES OF HEALTH INVESTIGATION, 14(1), 91–96. https://doi.org/10.21270/archi.v14i1.6446